Model |
11235BBC86(3') |
BSC-4FA2(4') |
11236BBC86(6') |
External Size(W*D*H) |
1087*775*2265mm |
1383×775×2295mm |
1873*775*2295mm |
Internal Size(W*D*H) |
910*600*660mm |
1210×660×660mm |
1700*600*660 mm |
Work Surface Height |
750mm |
Max Opening |
550mm (22’’) |
Tested Opening |
Safety height 200mm (8’’) |
Airflow Volume |
Inflow |
349m3/h(205cfm) |
465m3/h(275cfm) |
649m3/h(382cfm) |
|
Down flow |
64%:638m3/h(375cfm) |
64%:868m3/h(498cfm) |
64%:1190m3/h(700cfm) |
|
Exhaust |
36%:359m3/h(211cfm) |
36%:488m3/h(287cfm) |
36%:669m3/h(394cfm) |
Front Window |
Motorized. Two-layer laminated toughened glass ≥5mm. Anti UV |
HEPA Filter |
Two, 99.999% efficiency at 0.3um, filter life indicator |
Filter Guard Type |
Aluminum alloy frame |
Noise |
NSF 49≤61 dB/EN 12469≤58 dB |
Illumination |
≥1000Lux |
Display |
LCD Display |
Waterproof Socket |
2, Total consumption:≤500W |
Tap |
Water tap*1, Gas tap *1 |
Ground Resistance |
≤ 0.10Ω |
Power Supply |
AC220/110V±10%, 50/60Hz ±1Hz, Full load Amps: 9A, BTU/Hr: 1689 |
Consumption |
300W |
400W |
500W |
Alarm |
Abnormal airflow velocity; Filter change; Front Window over height |
Motor |
One ECM motor, 110V& 220V acceptable, Speed adjustable, |
high efficiency and low power consumption |
Material |
Working area: 304stainless steel/ Frame and Decorative Plate: cold-rolled steel |
with anti-bacteria power coating |
UV Lamp |
18W |
30W |
40W |
|
Germicidal UV lamp, Emission of 253.7 nanometers for most efficient decontamination |
Standard Accessory |
Fluorescent lamp, UV lamp *2, Base stand, SS water & gas taps, Drain valve, |
Waterproof sockets, Stainless Steel I.V. Bar with 10 hooks |
Optional Accessory |
Armrest, Airflow tester, Formalin fumigation sterilizer,Infrared sterilizer, |
Adjustable base stand, Remote control |
Package Size (W*D*H) |
1250*1080*1840mm |
1540*1080*1900mm |
2020*1080*1900mm |
Gross Weight |
270kg |
334kg |
363kg |