Characteristics:
l Sterilizing course automatic sterilization controlled,easy to operate.
l With drying function,suitable dressing drying.
l With over-temperature,over-pressure auto-protect device.
l The door opening mechanism cannot be operated until the pressure in the chamber is reduced to 0.027MPa.It can't be start on if the door doesn't close well.
l The safety value will be open automatically when the inner pressure over 0.24MPa,and the steam be exhausting to the water tank.
Specifications:Model | WS-150YDC | WS-200YDC | WS-280YDC | WS-400YDC |
Sterilizing chamber | 150L | 200L | 280L | 400L |
Volume | φ440×850mm | φ515×1000mm | φ600×1000mm | φ700×1100mm |
working pressure | 0.22MPa | |||
working temperature | 134℃ | |||
Adjustment of temperature | 40℃- 134℃ | |||
Time of sterilization | 0-99 min | |||
Time of drying | 0-99 min | |||
Heat average | ≤±2℃ | |||
Power | 9KW/380V 50Hz |
9KW/380V 50Hz |
12KW/380V 50Hz |
18KW/380V 50Hz |
Dimension | 1600×050×2100 | 1400×670×1650 | 1400×770×1780 | 1430×880×1830 |
Transport dimension | 1550×750×1850 | 1560×820×1850 | 1680×920×2100 | 1600×1050×2100 |
G.W/N.W | 320/240 KG | 350/260 KG | 465/365 KG | 530/420 KG |